VAC-V2 GAS PERMEABILITY TESTER – PERMEATION INSTRUMENT
 
VAC-V2 Gas Permeability Tester – Permeation Instrument

VAC-V2 Gas Permeability Tester – Permeation Instrument

ISO 2556, ISO 15105-1, ASTM D1434, GB 1038 JIS K7126-A, YBB 00082003

VAC-V2 Gas permeability tester is applicable in films and sheets test of gas transmission rate, solubility coefficient, diffusion coefficient and permeability coefficient under different temperatures.

Characteristics
  • Manometric principle, compatible with GB, ISO, ASTM standards.
  • Three independent test chambers, can display relative data, average data and statistic data.
  • World-famous vacuum sensor of highly precision ensuring accurate test data.
  • Temperature control device enables permeability test under any temperature between 5~75°C.
  • World-famous brand elements with stable and reliable performance.
  • Optional dual test modes: Conventional and non-standard.
  • Super powerful software support, automatically calculating solubility, permeability and diffusion coefficients. EXCEL format conversion, graph display.
  • World original design. Curvefitting Methodology for gas transmission rate, diffusion coefficient, solubility coefficient and permeability coefficient at any temperature.
  • Can easily get gas transmission rate, diffusion coefficient, solubility coefficient and permeability coefficient of the materials under temperatures between -273℃~200℃.
  • Intelligent parameter setting and test operation; Automatization of the whole test process.
  • Super reliable performance, nonstop operation, and improved test efficiency.
  • Satisfactory data repeatability and small margin for operating error.
  • Blank test with little systematic error; General performance exceeds similar products throughout the world.
  • Calibration of reference film, easy and convenient to operate.
Technical specifications of the VAC-V2 Gas Permeability Tester – Permeation Instrument
Test rage 0.05 ~ 50,000 cm³/m²·24h ·0.1MPa(normal)
upper limit not less than 500,000 cm³/m²·24h ·0.1MPa(extension volume)
Note:Normal and extension volume can be selected using extension vacuum accessory.
Temperature control range 5~75℃
Temperature control accuracy ±0.1℃
Vacuum accuracy 0.1Pa
Vacuum degree in test chamber <20Pa
Tested gas O2, N2, CO2, etc.
Test pressure –0.1MPa ~ +0.1MPa
Pressure of gas source 0.4MPa≤P≤0.6MPa
Specimen size Φ97mm
Transmission area 38.46cm² (diameter 70mm)
Dimension 760mm(L)×575mm(B)×450mm(H)
Power AC 220V 50Hz / 60Hz
Weight 88kg
Configuration of the VAC-V2 Gas Permeability Tester – Permeation Instrument
Standard Mainframe, software, extension vacuum accessory, sample cutter, vacuum pump, temperature controller.
Optional Sample cutter blades, vacuum grease, oil for vacuum pump, fast quantitative filter paper.
Note:Users provide test gas for themselves.


Oxygen / Water Vapor / Gas Permeability Tester and Permeation Instruments for Films and Finished Packages

For further information or to place an order for the VAC-V2 Gas Permeability Tester – Permeation Instrument, please CLICK HERE, call us Toll-Free at: 1.877.884.TEST, or click on the "HOW TO ORDER" Button