| Interferometer | Cube-corner Michelson interferometer |
| Beam splitter | Multilayer Ge coated KBr |
| Detector | High sensitivity pyroelectric module (standard) | Temperature stabilized, MCT detector (optional) |
| IR Source | High intensity, long lifetime, air-cooled IR source |
| Wavenumber Range | 7800cm-1~350cm-1 |
| Resolution | 0.85 cm-1 |
| Signal to noise ratio | Better than 20,000:1 (RMS value, at 2100cm-1 ~ 2000cm-1 or 2100cm-1 ~ 2200cm-1, resolution: 4cm-1, detector: pyroelectric, 1 minute data collection) |
| Wavenumber Accuracy | ±0.01 cm-1 |
| Scanning Speed | Microprocessor control, different scanning speed selectable. |
| Software | MainFTOS Suite software workstation, compatible to all version Windows OS | Various specialized functional software modules (optional) |
| Communication | Ethernet interface & WIFI wireless communication |
| Data Output | Standard data format, report generation and output |
| Status Diagnosis | Power on self-check, real-time temperature and humidity monitoring and reminders |
| Certification | FDA 21 CFR Part11 compliance, IQ/OQ/PQ (Optional) |
| Environment Conditions | Temperature: 10℃~30℃, Humidity: less than 60% |
| Power Supply | Universal between 110V – 220V |
| Dimensions & Weight | 490 x 420 x 240mm, 23.2kg |
| Accessories | Transmission sample holder (Standard) | Optional accessories such as gas cell, liquid cell, Defused/Specular Reflection, single/multi reflection ATR, etc. |