The OmniTest Universal Hardness Tester is designed to meet the increasing demand of versatility for the laboratory or quality control station. The unique design and advanced features eliminate operator influence during indentation measurement, thus assuring the highest repeatability. The combination of cutting edge technology and ease-of-use make the OmniTest with fully automatic image evaluation the obvious replacements for traditional mat-screen testers.
The OmniTest offers a load range between 1 and 250kgf.
Standardized procedure |
Rockwell |
DIN EN ISO 6508, ASTM E 18 |
Vickers |
DIN EN ISO 6507, ASTM E-384 |
|
Brinell |
DIN EN ISO 6506, ASTM E 10 |
|
Test loads |
Rockwell and Superficial Rockwell Preload |
3-10 kgf Load: 15, 30, 45, 60, 100, 150 kgf |
Vickers |
1, 2, 3, 4, 5, 10, 20, 30, 40, 50, 60, 80, 100, 120 kgf |
|
Brinell |
1, 2.5, 5, 6.25, 10, 15.625, 25, 30, 31.25, 62.5, 100, 125, 187.5, 250 kgf |
|
Test Range |
Depth Reading Methods |
Rockwell and Superficial Rockwell: Full range with 0.1HR resolution |
Optical Methods |
33x Lens measures indentation diameters: 0.66mm – 1.50mm |
|
133x Lens measures indentation diameters: 0.26mm – 0.66mm |
||
400x Lens measures indentation diameters: 0.03mm – 0.26mm |
||
(The minimum diameter stated for each lens reflects the minimum diameter that can be measured while in full compliance with all testing standards requirement for resolution. Smaller diameters can be measured but with some loss of accuracy) |
||
Indenters |
Rockwell Conical diamond indenter 120° |
|
Vickers Pyramidal indenter 136° |
||
Brinell Ball indenter 1mm, 2.5mm, 5mm, 10mm |
||
Ball indenter 1/16”, 1/8”, 1⁄4”, 1⁄2” (on request) |
||
Optical testing device: From 35-400x with interchangeable objectives (400x on request) LED lighting source |
||
Dimensions |
Max measurable height 290mm (11.4”) |
|
Max measurable depth 250mm (9.8”) |
||
Weight 250kg (550lb) |
||
Test load application |
DC motor via closed loop by means of force transducer |
|
Test load selection by LCD screen |
||
Automatic rotation to indenter/objective |
||
Display of Results |
Numeric on touch screen LCD (with automatic storage) Graphic capability for result analysis with variable scale |
|
Data Transfer, Connections, and power supply |
Interface RS232 and USB standard |
|
Standard power supply is 230/110V, 50/60 Hz. Other power supplies on request CE conformity |
1-877-884-8378