Specifications of UV-2100 UV/VIS Spectrophotometer | |
---|---|
Wavelength range | 190-1100nm |
Spectral Bandwidth | 0.1, 0.2, 0.5, 1.0, 2.0nm |
Wavelength accuracy | ±0.3nm(0.15nm if required) |
Wavelength Reproducibility | 0.15nm |
Photometric Accuracy | ±0.3%T(0-100%T), ±0.002A(0-0.5A), ±0.004A(0.5-1A) |
Photometric Reproducibility | 0.15%T |
Working mode | T, A(-3~4A), C, E |
Stray Light | ≤0.05%T(NaI, at 220nm) |
Baseline Flatness | ±0.001A |
Stability | 0.0004A/h (at 500nm, after warming up) |
Detector | Photomultiplier |
Light source | Tungsten halogen lamp and deuterium lamp |
Power | AC:230V/50Hz, 110V/60Hz, 400W |
Dimensions | 670×470×210mm |
Weight | 45kg |