Guarded Hot Plate Thermal Conductivity Tester

Request A Quote
Standards

Guarded Hot Plate Thermal Conductivity Tester - QualiGTX™ GHP

Guarded Hot Plate Apparatus for Insulation Boards and Mortars, 300 x 300 mm Specimen

The QualiGTX™ GHP Guarded Hot Plate Thermal Conductivity Tester measures the lambda value of insulation materials by the steady-state guarded hot plate method, covering 0.010 to 3.000 W/(m·K) on 300 × 300 mm specimens. Thermal conductivity is the number that decides whether an insulation product meets a building energy requirement, and the guarded hot plate is the primary method against which comparative instruments are calibrated. Test accuracy is held to within 2% with repeatability under 1%.

The instrument suits foam concrete, insulation mortars, and similar cast or formed materials, tested as standard specimens produced in a matching 300 × 300 mm mold. Specimen loading and unloading is automated with uniform clamping pressure, which matters more for rigid materials than it first appears: contact resistance between the plates and a slightly uneven specimen face is one of the larger error sources in the method, and consistent clamping keeps it constant from run to run.

Cold plate temperature is settable from 5 to 50 °C and hot plate temperature from ambient to 80 °C, so a laboratory can test at the mean temperature its specification calls for rather than at whatever the apparatus defaults to. Specimen thickness spans 10 to 45 mm with 25 mm preferred. The unit draws 2 kW from a single-phase supply, weighs 215 kg, and stands 1600 mm tall on an 800 × 600 mm footprint. Operation is specified for a 10 to 40 °C environment, with an air-conditioned laboratory recommended.