Low-Energy Charpy V-Notch Specimens SRM must be impact tested (fractured) consecutively within the same timeframe (not on separate days) at 21°C ± 1°C. Test results and digital photographs of the fractured samples should not be emailed for evaluation. A compliant machine will produce an average value within 1.4 J or 5% of the certified energy value, whichever is greater.
Unit of Issue: Set
Unit Description: Each unit contains five Charpy V-notch specimens required for one in-house verification (self-verification) of a machine equipped with a 2 mm striker. The specimens are manufactured from 4340 alloy steel.
Lot Number: LL-198
SKU: 2197
Validity: Indefinite
Safety Information
These SRMs are designed for the in-house indirect verification of Charpy testing machines equipped with 2 mm strikers, in compliance with the current ASTM E23 and ISO 148-2 standards. Testing for SRM 2197 must be conducted at 21°C ± 1°C.