Compost Degradation Tester QualiCDT™ 180 Auto Features
Precise Control of Experimental Conditions
- Independent CO2 Sensors: Incorporates 19 independent imported infrared carbon dioxide sensors ensuring accurate and independent measurement results.
- Temperature and Humidity Control: Each chamber features independent temperature and humidity sensors, with PID fuzzy algorithm intelligent control and condensed water return technology to maintain set conditions during tests.
- Automated Humidity Monitoring: Includes a built-in water tank with real-time humidity monitoring, automatically replenishing water to meet standard water content requirements for the inoculum.
- Mixing Systems: Features 18 independent mixing systems with a three-dimensional fence design. Allows customizable mixing periods, times, and stirring speeds, including forward and reverse stirring, effectively preventing inoculum hardening during the test cycle.
Technological Advancements
- Intelligent Air Tightness Monitoring: Equipped with high-precision flowmeters and real-time monitoring of system sealing performance. Includes a suspected leakage alarm function to ensure adequate aeration conditions in each composting container.
- Power Failure Support: Test function resumes seamlessly after accidental power failures.
- Flexible Channel Configuration: Channels can be freely combined with customizable blank, reference, and sample modes for independent testing.
- Real-Time Data Display: Provides real-time display of carbon dioxide quality curves and biodegradation rate curves.
- Remote Monitoring: Supports remote monitoring of the experimental process via mobile phones, iPads, and PCs.
Result Validity Analysis
- Validity Judgment Module: Software includes a module to assess result validity, displaying the biodegradation rate of the reference material after 45 days, carbon dioxide produced by the blank container within 10 days before the experiment relative to the inoculum's volatile solids, and the relative deviation of biodegradation rates across 18 channels.
- Comprehensive Reporting: Report templates include parameters such as the average carbon dioxide value of the blank group, theoretical and generated carbon dioxide, biological decomposition rates, relative decomposition rates, and time.
Personalized Customization
- Tail Gas Analysis Module: Customizable to meet various scientific research needs, including methane, ammonia, nitrogen dioxide, hydrogen sulfide, sulfur dioxide, odor, VOC, and other gas characteristic parameters.
- Station Customization: Supports customization for 12/18/24/30 stations and data mean mode preferences.
Intelligent Operating System and Global Certification
- Flexible Parameter Setting: Designed according to GMP appendix computer system guidelines, allowing flexible setting of test process parameters to meet different standard requirements.
- Audit Trail Function: Includes log trace and test trace functions for data traceability. Supports setting different levels of system operation authority.
- Customizable Test Reports: Allows personalized test reports, supports data output in multiple formats, electronic signatures, and online submission of audit reports.
Laboratory Intelligent IoT Platform
- IoT Connectivity: Connects to an IoT platform for networked digital management.
- Remote Access: Allows remote login to the IoT platform for managing experimental data, diagnosing, and resolving faults.
- Resource Accessibility: Customers can independently download necessary instrument information, documents, and operation videos from the platform.