Theory and Method
Kinematic viscosity describes how quickly a liquid flows under gravity. Capillary viscometry measures it by timing a fixed volume of sample as it drains between two calibrated marks in a narrow glass tube. The Hagen-Poiseuille relationship explains the result: flow time grows in direct proportion to kinematic viscosity, which equals dynamic viscosity divided by density. Thick oils drain slowly, and thin fuels drain quickly.
Each viscometer tube carries a calibration constant, C, determined against reference oils. The instrument multiplies the measured flow time, t, by that constant to calculate kinematic viscosity (ν = C × t), reported in mm²/s (equal to centistokes). Multiplying the result by sample density gives dynamic viscosity. Because viscosity shifts sharply with temperature, the method demands a thermally stable bath, which explains the ±0.01°C tolerance.
The QualiAKVT™-D445 automates each manual step of this procedure. The PID controller and PT100 probe hold the bath at setpoint, while the infrared photoelectric sensor detects the liquid and records flow time electronically. Light-blocking samples work the same way, because the sensor tracks the liquid itself instead of relying on operator observation. After timing, the instrument runs a wash sequence that prepares the tube for the next sample.